Anaesthesia Machine

Anaesthesia Machine
Product Detailed
1,Anaesthesia Machine: MT02002004 2,CE& ISO Approval 3,OEM&ODM Available 4,Good Quality & Competitive Price

 Anaesthesia Machine

 Anaesthesia,Medical Device

Specifications

Gas supply:O2/N2O/Air 0.27~0.55MPa

Flow adjustment:O2/N2O 0.05~10L/min

                Air 0.2~12L/min

Flow djustment:O2+NO or O2+Air

O2 flush:35~75L/min

Build-in Safety device:

 Oxyge ratio control(ORC)guarantee minimum 25% percentage of oxygen

 Oxygen deficiency alarm when O2≤0.2Mpa, audible alarm occurs and lasts for at least 7 seconds

 N2O cut-off when O2≤0.1Mpa,N2O cut-off is activated

 Backup power when AC power fails, the power switch over to internal battery which will run the machine for approximately four hours

Vaporizer:

 ZFG-B1 output concentration:0.5%~5% with T.P.F compensations

Draeger LC output concentration 0.2%~6%

Anaesthesia circuit:

 Relief valve:0.5~7kPa

 CO2 absorber volume:≥1.5L

Leakage:≤175ml/min

Anaesthesia ventilator:

 Ventilation Modes:Manual,Spontaneous, Mechanical CMV,etc.

 I:E ratio 1:1~1:9.9(Reversed I:E optional)

 Ventilation Frequency:6~40 cycles/min,(4~60 cycles/min optional)

 Tidal Volume 100~1400ml(standard),20~1400ml(optional)

 PEEP:0~2kPa

Monitoring:Exhaled tidal volume,Minute volume, airway pressure, ventilation frequency, Inspired oxyen concentration.

Alarm:Visual and Audible alarm for O2 supply low, Paw low, Paw high,MV low MV high, Apnea, O2% low,O2% high, battery low, batter insufficient Power filure.